Comparison of the ocular response analyzer and the Goldmann applanation tonometer for measuring intraocular pressure after penetrating keratoplasty


1 Department of Ophthalmology, Menoufia University, Menoufia, Egypt
2 Department of Ophthalmology, Memorial Institute of Ophthalmology, Giza, Egypt

Correspondence Address:
Eslam A. M. Hassan
Home 42, 5th District, New Cairo, Cairo 35611
Egypt
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/mmj.mmj_28_19

Rights and Permissions

Objectives To compare intraocular pressure (IOP) readings measured by the ocular response analyzer (ORA) with those measured by the Goldmann applanation tonometer (GAT) following penetrating keratoplasty (PKP) and evaluate the influence of biomechanical properties of the grafts on IOP measurements. Background The ORA (Reichert Technologies) is a device that shows the biomechanical properties of the cornea. It reflects certain biomechanical properties of the cornea such as corneal hysteresis (CH) and corneal resistance factor (CRF). Patients and methods IOP using the GAT [IOP (GAT)], CH, CRF, Goldmann-correlated IOP (IOPg), cornea-compensated IOP (IOPcc) using the ORA, and central graft thickness (CGT) were measured in 30 eyes undergoing PKP. Bland–Altman plots were used to evaluate agreement between the tonometers. The correlation between refraction, graft curvature, astigmatism, CH, CRF, and CGT with IOP readings was investigated using multivariate regression analysis. Results The mean CGT, CH, and CRF were 532.43 ± 30 μm, 8.52 ± 1.81 mmHg, and 8.56 ± 1.59 mmHg, respectively. Mean IOP (GAT), IOPg, and IOPcc were 11.88 ± 3.66, 14.64 ± 4.08, and 17.27 ± 4.60 mmHg, respectively. The 95% limit of agreement between IOP (GAT) and IOPg ranged from 2.39 to 3.29 mmHg, and for IOPcc and IOP (GAT) it ranged from 4.7 to 6.2 mmHg. No significant association was found between CGT and IOP readings. According to IOP (GAT), there was significant negative association with CH. According to IOPcc, there was significant negative association with CH and mean keratometry. According to IOPg, there was significant positive association with CRF. Conclusion Following PKP, graft biomechanics had more influence on IOP values than anatomical features. In comparison to the GAT, the ORA yielded higher IOP values.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed1084    
    Printed56    
    Emailed0    
    PDF Downloaded76    
    Comments [Add]    

Recommend this journal